DISSOLVE OXYGEN MEASUREMENT
DISSOLVE OXYGEN MEASUREMENT
Different methods are used for measurement of the
amount of dissolve oxygen present in a liquid. It measure the dissolve oxygen
in mg/L, partial pressure or a % of saturation or ppb.
There are three
techniques for DO measurement
1) Colorimetric
method
2) Titration
3) Sensor
and meter
· Colorimetric
method: It is very simple method mainly used for
determination of DO in water sample. Based on the reagent used, this method is
divided in two types.
a) Indigo
carmine method: In this method, reduced indigo carmine dye added into the sample
that react with the presence of dissolve oxygen and undergo color changes (blue
color), and change in color is directly proportional to DO. This color change
compare with color comparator and determine the DO. LOD of this method is 0 to
50 ppm.
b) Rhodazine
D method: This method is used to determine the dissolve oxygen in very low
concentration in ppb. In this method, rhodazine D reagent react with dissolve
oxygen and to form deep rose color which is directly proportional to DO, and
determine by using color comparator.
· Titration
method: Traditional method is the winkler titration method,
but have under question of human error. This test is mainly use for water
quality study.
Principle:
In this method manganese hydroxide {Mn(OH)2} is added to the sample
which react with dissolve oxygen molecules in the sample and produce brown
precipitate of manganic oxide {MnO(OH)2}, Which react with
concentrated sulfuric acid to liberate nascent oxygen. Potassium iodide (KI) is
added which react with nascent oxygen and convert to iodine (I2) molecules
that is then titrated against the sodium thiosulphate using starch as an
indicator. Thus, the amount of sodium thiosulphate consumed is directly proportional
to dissolve oxygen concentration.
· Sensor
and meter: Two type sensors are available (1) Optical and (2)
Electrochemical (ampherometric /Clark type).
1)
Optical
sensor:
Working
Principle: Detection based on fluorescent quenching, fluorescent lifetime
detection (phase shift) and fluorescent intensity detection (through
photodiode) as O2 affect both.
It
measure the interaction between O2 and luminescent dye, when this dye expose to
blue light electron get the energy go to excited stage and come back to normal
stage post releasing the energy. When O2 molecules are present, it interact
with the dye and alter the return wavelength. The measured effect is inversely
proportional to the partial pressure of O2. The intensity of lifetime of
returned luminescence is measured by photodetector and can be used to calculate
the DO concentration.
Advantages:
o
Fast response time,
o
Doesn’t require calibration, membrane,
filling solution/electrolyte, stirring, warm up time
o
Not affected by hydrogen sulfide or other
gases that can permeate the electrochemical DO membrane
Disadvantages:
o
Susceptible to external temperature,
pressure and salinity
o
Attenuation of light source
o
Drift due to degradation and leaching of
the dye
2) Electrochemical sensor:
Electrochemical sensor further break down in two types a) Polarography sensor
and b) Galvanic sensor
a) Polarography sensor:
Polarography sensor break down in two types, polarography and pulse
polarography. In polarography sample needed to stir while in pulse polarography
there is no need to stir the sample. Both the sensor have two polarized
electrodes, an anode made up of silver-silver chloride and a cathode made up of
gold/platinum in an electrolyte solution and are isolated by thin semipermeable
membrane. At a time of measurement, dissolve oxygen diffuse via membrane at a
rate of partial pressure of oxygen. This oxygen undergoes the reduction at
cathode and produce an electric current; this current is carried out by ions in
the electrolyte from cathode to anode. This current is proportional to partial
pressure of oxygen.
b) Galvanic sensor: This
sensor having two electrode, cathode and anode immersed in electrolyte. The
oxygen from the sample diffuse to the cathode and reduced to hydroxyl ion. The
hydroxyl ion moves to anode where it oxidized, and generate a current. This
current is directly proportional to concentration of oxygen.
Digital
sensor: Digital sensor available in market with a memory
storage device between cable and probe so once it calibrated can be used
directly to the other fermenter without recalibration.
Analog
sensor: There is no memory device in this sensor so needed to
recalibrate whenever cable changed or dismissed from the cable.
DO
Probe Calibration: Calibration can be carried out at two
points,
(1) zero percent calibration and (2)100% calibration.
●Zero percent calibration
can be performed by i) using zero gel/saturated solution of sodium sulfite, ii) by
purging the nitrogen into the sample, iii)by dismantling the probe from the cable
or iv) at sterilization hold time.
●100% calibration can be
performed i) in air in the open environment, ii) in fermentation post sterilization
with working aeration and agitation.
In fermentation process probe calibration should be carried out after sterilization of media because during sterilization there is a chances of tear and wear of membrane and there is a chances of sticking of degraded media component to the membrane which gives false result.
DO
Probe Polarization: For new DO probe polarization six hours is required, For routine if probe undergoes depolarization for short time like 5-10 minutes than
three time of total depolarization time required for activation of it.
I.e. If probe goes to depolarization for 10min than it required to keep for polarization (10*3=30min) at least 30min.
DO
Probe Storage: For long-term storage, stored in dry condition in
which remove electrolyte from membrane give water wash to probe and stored in
dry condition. For short-term storage stored in water.
Membrane:
Teflon
membrane (T type membrane available having three layers Teflon-ss-teflon,
useful for longtime process), polythene membrane, SS membrane (S type membrane
available having two layers ss-teflon, useful for short time process.
Factor
Affecting to DO:
Temperature – Inversely
proportional, as temperature increases DO decreases. At ≥ 100 ͦc it shows zero.
Pressure -Directly
proportional, as pressure increases DO increases.
Salinity - Inversely
proportional, as salinity increases DO will decreases.
Aeration -Directly
proportional, as aeration increases DO will increases.
Agitation- Directly
proportional, as agitation increases DO will increase.
Comments
Post a Comment